Skroot Contact us
SUPPORT

Here to help you implement, run and analyze

From first setup to reading your results, our engineers work alongside you. Manuals and FAQs are below, and a direct line is one email away.

EMAIL SUPPORT
support@skrootlab.com

Setup questions, vessel fit, analysis help or a run you would like a second opinion on. Include your device ID and we reply within one business day.

Manuals and downloads

Sorted by product.

RESEARCH USE ONLY

SMART™ Explorer

Product page →
PDF · v1.4 SMART™ Explorer User Guide Setup, calibration, running and exporting, with a quick-start checklist for your first run. Download → PDF SMART™ Sensor Application Guide Step-by-step instructions for applying a SMART™ sensor. Download →
VIDEO TUTORIALS
Running the SMART™ Explorer
Applying a SMART™ Sensor
Applying a SMART™ Sensor to Non-Scratch Surfaces
DEVELOPED WITH IBI SCIENTIFIC

SMART-Cell Platform

Product page →
PDF · FROM IBI SCIENTIFIC SMART-Cell Reader Operation Manual Setup, priming and operation of the SMART-Cell Reader, hosted by IBI Scientific. Download at IBI → ORDERING AND WARRANTY IBI Scientific Orders, replacement SMART-Cell Sensors, shipping and warranty for the SMART-Cell Reader. Contact IBI →
VIDEO TUTORIALS
SMART-Cell Dashboard Tutorial
FOR INTEGRATION PARTNERS

OEM Kits

Product page →
SHARED WITH PARTNERS Integration documentation Hardware and software documentation for the SIB board, antenna and Skroot software is shared directly with integration partners. Contact us → ORDERING OEM Hardware Kit and OEM SMART™ Sensor Kit Sold separately. SMART™ sensors have a minimum order of 10. OEM purchases are limited to 10 readers and 500 SMART™ sensors per customer per year. Request a quote →
VIDEO TUTORIALS
Applying a SMART™ Sensor
Applying a SMART™ Sensor to Non-Scratch Surfaces
ALL PRODUCTS

Frequently asked

Getting started, sensors and sterilization, your data, and ordering.

GETTING STARTED
How do I run my first culture?

Prepare the culture and make sure the sensor is submerged. Calibrate the reader with the coupler empty. Center the sensor over the coupler, parallel and close, then scan and confirm a clear resonance peak. Start the run with timed scans or single scans on demand. Full detail is in the SMART™ Explorer User Guide.

Does the SMART™ Explorer need an internet connection?

No. The SMART™ Explorer works fully offline and exports CSV over USB. Wi-Fi is only needed for the optional Skroot Data Portal, and the Explorer keeps recording if the connection drops.

SENSORS, CLEANING AND STERILIZATION
Are SMART™ sensors single use?

Yes. SMART™ stands for Single-use, Metabolite Absorbing, Resonant Transducer. Use a new sensor for each culture run.

Can SMART™ sensors be sterilized?

SMART™ sensors have been tested for compatibility with gamma irradiation (25–40 kGy nominal dose), steam autoclaving and UV exposure after application to a vessel wall. The release liners were outside the scope of this compatibility testing.

Before sterilizing the vessel, follow the sensor application instructions and confirm that both release liners have been removed. The top liner is blue with the Skroot logo, and the bottom liner is red.

How do I clean the reader and touchscreen?

Clean and disinfect the reader and touchscreen with standard laboratory cleaners. Both stay outside the vessel and never contact the culture.

How do I restock SMART™ sensors?

SMART™ sensors can be restocked with the OEM SMART™ Sensor Kit. Email us with the quantity you need and we will send a quote.

YOUR DATA
What does the SMART™ Growth Index measure, and what does it not measure?

The SMART™ Growth Index follows growth-associated metabolites that the sensor absorbs from the medium as cells grow. It responds to a broad class of compounds rather than one specific molecule. It does not directly measure cell count or viability, so pair it with your usual reference samples when you need those numbers.

How do I get my data into Excel, Python or MATLAB?

Use Export Run on the touchscreen to save a CSV to USB. With the Data Portal, download CSV, XLS, PNG or PDF from any chart, mid-run.

Where is dashboard data stored?

The Skroot Data Portal and the SMART-Cell Dashboard are securely hosted on Amazon Web Services (AWS).

Can you help me analyze my results?

Yes. Export the run and email it with the device and run IDs. We will walk through the curve with you and help line the SMART™ Growth Index up with your reference measurements.